How is detection limit of fluorescence calculated?

The linear equation was found to be y=31.438x-106.557(R=0.998), where y is the fluorescent intensity data at 562 nm measured at a given Hg2+ concentration and x represents the concentration of Hg2+ added. So the detection limit for Hg2+ was calculated to be 25nM(Detection limit=3σ/K=0.801/31.438×10-6 M ≈ 25nM).

What is the range of fluorescence spectroscopy?

Fluorescence is used mainly for measuring compounds in solution. In fluorescence spectroscopy, a beam with a wavelength varying between 180 and ∼800 nm passes through a solution in a cuvette.

What is the detection of fluorescence?

In some cases, when the light energy is absorbed by a molecule, it raises some of the electrons to an excited state. When these electrons return to the ground state and light is emitted, the process is referred to as fluorescence. Fluorescence detectors rely on this molecular property for detection.

Which detector is used in fluorescence spectrophotometer?

A fluorometer or fluorimeter is a device used to measure parameters of visible spectrum fluorescence: its intensity and wavelength distribution of emission spectrum after excitation by a certain spectrum of light. These parameters are used to identify the presence and the amount of specific molecules in a medium.

What is detection limit of a spectroscopy?

The detection limit is (informally) the lowest concentration of the analyte that can be reliably detected, and is a reflection of the precision of the instrumental response obtained by the method when the concentration of the analyte is zero.

Why does fluorescence have a lower detection limit than absorbance?

The reason why fluorescence is more sensitive than UV-Vis absorption is that they are measured in different ways. Absorbance is measured as the difference in intensity between light passing through the reference and the sample, whereas fluorescence is measured directly without any reference beam.

How is RFU measured?

The terms “relative fluorescence units” (RFU) and “RFU peak” refer to measurements in electrophoresis methods, such as for DNA analysis. A “relative fluorescence unit” is a unit of measurement used in analysis which employs fluorescence detection. The data can be normalized to DNA input or additional normalizing genes.

How accurate is fluorescence spectroscopy?

Fluorescence spectroscopy is a relevant analytical technique because of its extreme sensitivity and excellent specificity. Indeed, spectrofluorometric methods can detect concentration of component as low as one part in 1010, with its sensitivity 1000 times greater than that of most spectrophotometric methods.

What is the use of fluorescence detection?

Fluorescence detection is generally used for analysis when sensitivity and selectivity are required, especially when the analyte has little or no UV absorbance and can be derivatized to produce fluorescence.

Why are detectors placed at 90 degrees?

The sample emits a wavelength, which travels to the detector. The detector is usually set at a 90-degree angle to the light source to avoid any interference from the transmitted excitation light. Photons emitted hit a photo detector.

Which detector is commonly used in fluorescence?

Fluorescence (FL) detector A UV/UV-VIS detector monitors the absorption of light with a specified wavelength.

How is detection limit determined?

Based on visual evaluation: The detection limit is determined by the analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte can be quantified with acceptable accuracy and precision.

How are UV-Vis and fluorescence spectroscopy the same?

UV-Vis measures the absorption of light in this range, while fluorescence measures the light emitted by a sample in this range after absorbing light at a higher energy than it is emitting. 1.1 UV-Visible Spectroscopy

How is the absorption of UV Vis light measured?

UV-Visible absorption spectroscopy involves measuring the absorbance of light by a compound as a function of wavelength in the UV-visible range. When a molecule absorbs a photon of UV-Vis light, the molecule is excited from its ground state to an electronic excited state.

What are the components of a fluorescence detector?

All fluorescence instruments contain three basic items: a source of light, a sample holder and a detector. In addition, to be of analytical use, the wavelength of incident radiation needs to be selectable and the detector signal capable of precise manipulation and presentation (Figure 5).